Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9600 Schistosoma mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis

Lot 2422B-R (Expiry date: 25.02.2026)

 

Production date : 26.06.2024

Results of quality control 2 (after 6 months & rework), date: 23.12.2024

No serum blank (average) : 0.161

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.008 0.005 0.005 -0.003 0.000 0
Negative control serum
(N° 9600-06)
0.068 0.032 0.080 0.054 0.059 3
Weak positive serum
(N° 9600-07)
0.422 0.434 0.434 0.416 0.427 19
Positive control serum
(N° 9600-08)
2.217 2.387 2.247 2.284 2.284 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.284 Yes
Absorbance of weak positive serum 9600-07, as % of positive serum > 10 % 19 % Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.161 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch