#9600 Schistosoma mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis

Lot 2422B-R (Expiry date: 25.02.2026)

 

Production date : 26.06.2024

Results of quality control 1 (after production & rework), date: 05.09.2024

No serum blank (average) : 0.146

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.001 0.006 -0.003 -0.001 0.000 0
Negative control serum
(N° 9600-06)
0.069 0.073 0.051 0.075 0.067 3
Weak positive serum
(N° 9600-07)
0.500 0.539 0.482 0.468 0.497 20
Positive control serum
(N° 9600-08)
2.399 2.670 2.247 2.538 2.463 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.463 Yes
Absorbance of weak positive serum 9600-07, as % of positive serum > 10 % 20 % Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.146 Yes

Comments:

Good for use.