#9310 Echinococcus multilocularis (Em18) IgG ELISA
ELISA kit for human alveolar echinococcosis patient follow-up

Lot 2420F (Expiry date: 02.02.2026)

Production date : 03.06.2024

Results of quality control 1 (after production), date: 03.07.2024

No serum blank (average) : 0.113

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 -0.001 0.003 -0.001 0.000 0
Negative control serum
(N° 9310-06)
0.026 0.027 0.012 0.021 0.021 1
Weak positive serum
(N° 9310-07)
0.156 0.141 0.117 0.124 0.134 6
Positive control serum
(N° 9310-08)
2.523 2.695 2.222 2.277 2.429 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9310-08 > 1.200 2.429 Yes
Absorbance of weak positive serum 9310-07, as % of positive serum > 3 % 6 % Yes
Absorbance of negative serum 9310-06, as % of positive serum < 3 % 1 % Yes
Absorbance of no serum blank < 0.350 0.113 Yes

Comments:

Good for use.