#9310 Echinococcus multilocularis (Em18) IgG ELISA
ELISA kit for human alveolar echinococcosis patient follow-up

Lot 2420F-R (Expiry date: 02.02.2026)

Production date : 03.06.2024

Results of quality control 2 (after 6 months & rework), date: 29.11.2024

No serum blank (average) : 0.115

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.005 0.003 -0.001 0.003 0.000 0
Negative control serum
(N° 9310-06)
0.034 0.043 0.044 0.030 0.038 2
Weak positive serum
(N° 9310-07)
0.098 0.115 0.114 0.133 0.115 5
Positive control serum
(N° 9310-08)
2.141 2.422 2.237 2.330 2.283 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9310-08 > 1.200 2.283 Yes
Absorbance of weak positive serum 9310-07, as % of positive serum > 3 % 5 % Yes
Absorbance of negative serum 9310-06, as % of positive serum < 3 % 2 % Yes
Absorbance of no serum blank < 0.350 0.115 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland