#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2419S (Expiry date : 28.01.2026)
Production date : 29.05.2024
Results of quality control 1 (after
production), date: 03.06.2024
No serum blank (average) : 0.121
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
-0.003 |
0.006 |
-0.002 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.052 |
0.047 |
0.063 |
0.061 |
0.056 |
2 |
Weak positive serum
(N° 9450-07) |
0.695 |
0.700 |
0.715 |
0.775 |
0.722 |
29 |
Positive control serum
(N° 9450-08) |
2.378 |
2.370 |
2.366 |
2.673 |
2.447 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.447 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
29 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.121 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bâtiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|