#9450 Strongyloïdes ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis

Lot 2419S (Expiry date : 28.01.2026)

Production date : 29.05.2024

Results of quality control 1 (after production), date: 03.06.2024

No serum blank (average) : 0.121

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 -0.003 0.006 -0.002 0.000 0
Negative control serum
(N° 9450-06)
0.052 0.047 0.063 0.061 0.056 2
Weak positive serum
(N° 9450-07)
0.695 0.700 0.715 0.775 0.722 29
Positive control serum
(N° 9450-08)
2.378 2.370 2.366 2.673 2.447 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9450-08 > 1.200 2.447 Yes
Absorbance of weak positive serum 9450-07, as % of positive serum > 18 % 29 % Yes
Absorbance of negative serum 9450-06, as % of positive serum < 10 % 2 % Yes
Absorbance of no serum blank < 0.350 0.121 Yes

Comments:

Good for use.