#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2419S-R (Expiry date : 28.01.2026)
Production date : 29.05.2024
Results of quality control 1 (after
production & rework), date: 03.09.2024
No serum blank (average) : 0.134
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.002 |
-0.003 |
-0.001 |
0.001 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.106 |
0.103 |
0.120 |
0.137 |
0.117 |
4 |
Weak positive serum
(N° 9450-07) |
0.928 |
0.891 |
0.930 |
0.948 |
0.925 |
33 |
Positive control serum
(N° 9450-08) |
2.652 |
2.836 |
2.680 |
2.985 |
2.789 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.789 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
33 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.134 |
Yes |
Comments:
Good for use.