#9800
Anisakidae IgG ELISA
ELISA kit for the diagnosis of human anisakidosis
Lot 2418N (Expiry date: 06.01.2026)
Production date : 07.05.2024
Results of quality control 1 (after
production), date: 13.05.2024
No serum blank (average) : 0.123
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.003 |
0.003 |
-0.003 |
0.003 |
0.000 |
0 |
Negative control serum
(N° 9800-06) |
0.047 |
0.062 |
0.060 |
0.060 |
0.057 |
2 |
Weak positive serum
(N° 9800-07) |
0.351 |
0.361 |
0.345 |
0.367 |
0.356 |
15 |
Positive control serum
(N° 9800-08) |
2.211 |
2.479 |
2.256 |
2.545 |
2.372 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9800-08 |
> 1.200 |
2.372 |
Yes |
Absorbance of weak positive serum 9800-07, as % of positive serum |
> 9 % |
15 % |
Yes |
Absorbance of negative serum 9800-06, as % of positive serum |
< 8 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.123 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|