#9800
Anisakidae IgG ELISA
ELISA kit for the diagnosis of human anisakidosis
Lot 2418N-R (Expiry date: 06.01.2026)
Production date : 07.05.2024
Results of quality control 3 (after
12 months & rework), date: 02.05.2025
No serum blank (average) : 0.117
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
0.000 |
0.000 |
0.001 |
0.000 |
0 |
Negative control serum
(N° 9800-06) |
0.090 |
0.091 |
0.068 |
0.079 |
0.082 |
3 |
Weak positive serum
(N° 9800-07) |
0.403 |
0.430 |
0.354 |
0.378 |
0.392 |
16 |
Positive control serum
(N° 9800-08) |
2.320 |
2.569 |
2.204 |
2.402 |
2.374 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9800-08 |
> 1.200 |
2.374 |
Yes |
Absorbance of weak positive serum 9800-07, as % of positive serum |
> 9 % |
16 % |
Yes |
Absorbance of negative serum 9800-06, as % of positive serum |
< 8 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.117 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|