#6100 Aspergillus fumigatus IgG ELISA
ELISA kit for the diagnosis of human aspergillosis

Lot 2417A (Expiry date: 15.12.2025)

 

Production date : 16.04.2024

Results of quality control 1 (after production), date: 19.04.2024

No serum blank (average) : 0.117

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.005 -0.004 -0.001 -0.001 0.000 0
Negative control serum
(N° 61
00-06)
0.046 0.046 0.058 0.053 0.050 2
Weak positive serum
(N° 61
00-07)
0.444 0.453 0.507 0.509 0.478 19
Positive control serum
(N° 61
00-08)
2.435 2.507 2.467 2.548 2.489 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 2.489 Yes
Absorbance of weak positive serum 6100-07, as % of positive serum > 11 % 19 % Yes
Absorbance of negative serum 6100-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.117 Yes

Comments:

Good for use.