#6100 Aspergillus fumigatus IgG ELISA
ELISA kit for the diagnosis of human aspergillosis

Lot 2417A-R (Expiry date: 15.12.2025)

 

Production date : 16.04.2024

Results of quality control 4 (after 20 months & rework), date: 16.12.2025

No serum blank (average) : 0.122

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 -0.002 0.005 -0.001 0.000 0
Negative control serum
(N° 61
00-06)
0.052 0.045 0.053 0.046 0.049 2
Weak positive serum
(N° 61
00-07)
0.380 0.371 0.354 0.373 0.370 18
Positive control serum
(N° 61
00-08)
2.055 2.075 2.032 2.024 2.047 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 2.047 Yes
Absorbance of weak positive serum 6100-07, as % of positive serum > 11 % 18 % Yes
Absorbance of negative serum 6100-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.122 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland