Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#6100 Aspergillus fumigatus IgG ELISA
ELISA kit for the diagnosis of human aspergillosis

Lot 2417A-R (Expiry date: 15.12.2025)

 

Production date : 16.04.2024

Results of quality control 3 (after 12 months & rework), date: 11.04.2025

No serum blank (average) : 0.120

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.002 -0.003 -0.003 0.003 0.000 0
Negative control serum
(N° 61
00-06)
0.054 0.046 0.043 0.041 0.046 2
Weak positive serum
(N° 61
00-07)
0.403 0.398 0.421 0.390 0.403 18
Positive control serum
(N° 61
00-08)
2.190 2.288 2.299 2.144 2.231 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 2.231 Yes
Absorbance of weak positive serum 6100-07, as % of positive serum > 11 % 18 % Yes
Absorbance of negative serum 6100-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.120 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch