Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#6100 Aspergillus fumigatus IgG ELISA
ELISA kit for the diagnosis of human aspergillosis

Lot 2417A-R (Expiry date: 15.12.2025)

 

Production date : 16.04.2024

Results of quality control 2 (after 6 months & rework), date: 14.10.2024

No serum blank (average) : 0.128

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.011 0.005 -0.003 0.010 0.000 0
Negative control serum
(N° 61
00-06)
0.068 0.054 0.088 0.076 0.071 3
Weak positive serum
(N° 61
00-07)
0.589 0.567 0.626 0.621 0.601 22
Positive control serum
(N° 61
00-08)
2.540 2.542 2.683 2.961 2.681 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 2.681 Yes
Absorbance of weak positive serum 6100-07, as % of positive serum > 11 % 22 % Yes
Absorbance of negative serum 6100-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.128 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch