Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#6100 Aspergillus fumigatus IgG ELISA
ELISA kit for the diagnosis of human aspergillosis

Lot 2417A-R (Expiry date: 15.12.2025)

 

Production date : 16.04.2024

Results of quality control 1 (after production & rework), date: 05.09.2024

No serum blank (average) : 0.130

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.010 0.002 -0.003 0.012 0.000 0
Negative control serum
(N° 61
00-06)
0.044 0.075 0.108 0.074 0.075 3
Weak positive serum
(N° 61
00-07)
0.734 0.762 0.705 0.723 0.731 25
Positive control serum
(N° 61
00-08)
2.839 3.088 2.788 3.053 2.942 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 6100-08 > 1.200 2.942 Yes
Absorbance of weak positive serum 6100-07, as % of positive serum > 11 % 25 % Yes
Absorbance of negative serum 6100-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.130 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch