#9300 Echinococcus multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis

Lot 2416M (Expiry date: 15.12.2025)

 

Production date : 16.04.2024

Results of quality control 4 (after 20 months), date: 16.12.2025

No serum blank (average) : 0.125

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 0.003 -0.005 0.004 0.000 0
Negative control serum
(N° 9300-06)
0.014 0.019 0.027 0.018 0.020 1
Weak positive serum
(N° 9300-07)
0.563 0.511 0.545 0.547 0.542 24
Positive control serum
(N° 9300-08)
2.188 2.307 2.171 2.506 2.293 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.293 Yes
Absorbance of weak positive serum 9300-07, as % of positive serum > 14 % 24 % Yes
Absorbance of negative serum 9300-06, as % of positive serum < 6 % 1 % Yes
Absorbance of no serum blank < 0.350 0.125 Yes

Comments:

Good for use.