#9300
Echinococcus
multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis
Lot 2416M (Expiry date: 15.12.2025)
Production date : 16.04.2024
Results of quality control 4 (after
20 months), date: 16.12.2025
No serum blank (average) : 0.125
| Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.003 |
0.003 |
-0.005 |
0.004 |
0.000 |
0 |
Negative control serum
(N° 9300-06) |
0.014 |
0.019 |
0.027 |
0.018 |
0.020 |
1 |
Weak positive serum
(N° 9300-07) |
0.563 |
0.511 |
0.545 |
0.547 |
0.542 |
24 |
Positive control serum
(N° 9300-08) |
2.188 |
2.307 |
2.171 |
2.506 |
2.293 |
100 |
Validation criteria:
| |
Criterion |
Measure |
Valid |
| Absorbance of positive serum 9300-08 |
> 1.200 |
2.293 |
Yes |
| Absorbance of weak positive serum 9300-07, as % of positive serum |
> 14 % |
24 % |
Yes |
| Absorbance of negative serum 9300-06, as % of positive serum |
< 6 % |
1 % |
Yes |
| Absorbance of no serum blank |
< 0.350 |
0.125 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|