#9300 Echinococcus multilocularis (Em2-Em18) IgG ELISA
ELISA kit for the diagnosis of human alveolar echinococcosis

Lot 2416M (Expiry date: 15.12.2025)

 

Production date : 16.04.2024

Results of quality control 1 (after production), date: 19.04.2024

No serum blank (average) : 0.117

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.003 -0.004 0.001 0.001 0.000 0
Negative control serum
(N° 9300-06)
0.020 0.016 0.015 0.015 0.016 1
Weak positive serum
(N° 9300-07)
0.578 0.563 0.581 0.575 0.574 25
Positive control serum
(N° 9300-08)
2.152 2.302 2.214 2.339 2.252 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9300-08 > 1.200 2.252 Yes
Absorbance of weak positive serum 9300-07, as % of positive serum > 14 % 25 % Yes
Absorbance of negative serum 9300-06, as % of positive serum < 6 % 1 % Yes
Absorbance of no serum blank < 0.350 0.117 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland