#9750
Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis
Lot 2415R (Expiry date: 08.12.2025)
Production date : 09.04.2024
Results of quality control 3 (after
12 months), date: 04.04.2025
No serum blank (average) : 0.113
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.001 |
0.004 |
-0.004 |
0.001 |
0.000 |
0 |
Negative control serum
(N° 9750-06) |
0.025 |
0.031 |
0.034 |
0.032 |
0.030 |
1 |
Weak positive serum
(N° 9750-07) |
0.283 |
0.302 |
0.278 |
0.292 |
0.288 |
13 |
Positive control serum
(N° 9750-08) |
2.036 |
2.470 |
2.054 |
2.401 |
2.240 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9750-08 |
> 1.200 |
2.240 |
Yes |
Absorbance of weak positive serum 9750-07, as % of positive serum |
> 10 % |
13 % |
Yes |
Absorbance of negative serum 9750-06, as % of positive serum |
< 8 % |
1 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.113 |
Yes |
Comments:
Good for use.