Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9750 Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis

Lot 2415R (Expiry date: 08.12.2025)

 

Production date : 09.04.2024

Results of quality control 2 (after 6 months), date: 07.10.2024

No serum blank (average) : 0.116

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 0.006 -0.002 -0.003 0.000 0
Negative control serum
(N° 9750-06)
0.021 0.028 0.032 0.024 0.027 1
Weak positive serum
(N° 9750-07)
0.320 0.310 0.311 0.302 0.311 13
Positive control serum
(N° 9750-08)
2.193 2.327 2.327 2.384 2.308 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9750-08 > 1.200 2.308 Yes
Absorbance of weak positive serum 9750-07, as % of positive serum >  10 % 13 % Yes
Absorbance of negative serum 9750-06, as % of positive serum < 8 % 1 % Yes
Absorbance of no serum blank < 0.350 0.116 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch