#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2414D (Expiry date: 08.12.2025)
Production date : 09.04.2024
Results of quality control 3 (after
12 months), date: 04.04.2025
No serum blank (average) : 0.114
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
0.001 |
0.003 |
-0.001 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.059 |
0.073 |
0.074 |
0.068 |
0.068 |
3 |
Weak positive serum
(N° 9650-07) |
0.311 |
0.348 |
0.312 |
0.328 |
0.325 |
15 |
Positive control
serum
(N° 9650-08) |
1.968 |
2.295 |
2.024 |
2.274 |
2.140 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
2.140 |
Yes |
Absorbance of weak positive serum 9650-07, as % of positive serum |
> 8 % |
15 % |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 8 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.114 |
Yes |
Comments:
Good for use.