#9650 Fasciola hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 2414D (Expiry date: 08.12.2025)

 

Production date : 09.04.2024

Results of quality control 3 (after 12 months), date: 04.04.2025

No serum blank (average) : 0.114

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 0.001 0.003 -0.001 0.000 0
Negative control serum
(N° 965
0-06)
0.059 0.073 0.074 0.068 0.068 3
Weak positive serum
(N° 965
0-07)
0.311 0.348 0.312 0.328 0.325 15
Positive control serum
(N° 9650
-08)
1.968 2.295 2.024 2.274 2.140 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.140 Yes
Absorbance of weak positive serum 9650-07, as % of positive serum > 8 % 15 % Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.114 Yes

Comments:

Good for use.