#9650 Fasciola hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 2414D (Expiry date: 08.12.2025)

 

Production date : 09.04.2024

Results of quality control 2 (after 6 months), date: 07.10.2024

No serum blank (average) : 0.125

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.005 0.013 -0.006 -0.001 0.000 0
Negative control serum
(N° 965
0-06)
0.080 0.059 0.069 0.054 0.065 3
Weak positive serum
(N° 965
0-07)
0.358 0.386 0.326 0.335 0.351 16
Positive control serum
(N° 9650
-08)
2.175 2.390 2.110 2.228 2.226 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.226 Yes
Absorbance of weak positive serum 9650-07, as % of positive serum > 8 % 16 % Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.125 Yes

Comments:

Good for use.