#9650 Fasciola hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 2414D (Expiry date: 08.12.2025)

 

Production date : 09.04.2024

Results of quality control 1 (after production), date: 12.04.2024

No serum blank (average) : 0.140

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.001 0.005 -0.005 -0.001 0.000 0
Negative control serum
(N° 965
0-06)
0.061 0.065 0.080 0.067 0.068 3
Weak positive serum
(N° 965
0-07)
0.322 0.337 0.323 0.284 0.317 15
Positive control serum
(N° 9650
-08)
2.068 2.288 2.131 2.156 2.161 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.161 Yes
Absorbance of weak positive serum 9650-07, as % of positive serum > 8 % 15 % Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.140 Yes

Comments:

Good for use.