Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9550 Entamoeba histolytica IgG ELISA
ELISA kit for the diagnosis of human visceral amebiasis

Lot 2412H (Expiry date: 28.10.2025)

 

Production date : 29.02.2024

Results of quality control 1 (after production), date: 12.03.2024

No serum blank (average) : 0.155

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.016 0.032 -0.019 0.003 0.000 0
Negative control serum
(N° 9550-06)
0.103 0.176 0.087 0.133 0.124 5
Weak positive serum
(N° 9550-07)
0.746 0.778 0.795 0.874 0.798 32
Positive control serum
(N° 9550-08)
2.351 2.504 2.464 2.752 2.517 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9550-08 > 1.200 2.517 Yes
Absorbance of weak positive serum 9550-07, as % of positive serum >  17 % 32 % Yes
Absorbance of negative serum 9550-06, as % of positive serum < 12 % 5 % Yes
Absorbance of no serum blank < 0.350 0.155 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch