#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2411G (Expiry date: 28.10.2025)
Production date : 29.02.2024
Results of quality control 3 (after
12 months), date: 26.02.2025
No serum blank (average) : 0.134
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.017 |
-0.011 |
0.040 |
-0.013 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.027 |
0.020 |
0.034 |
0.034 |
0.028 |
1 |
Weak positive serum
(N° 9350-07) |
0.977 |
1.003 |
1.016 |
1.048 |
1.011 |
45 |
Positive control serum
(N° 9350-08) |
2.173 |
2.319 |
2.191 |
2.378 |
2.265 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.265 |
Yes |
Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
45 % |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
1 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.134 |
Yes |
Comments:
Good for use.