#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2411G (Expiry date: 28.10.2025)
Production date : 29.02.2024
Results of quality control 2 (after
6 months), date: 22.08.2024
No serum blank (average) : 0.110
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.001 |
-0.003 |
0.001 |
0.004 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.040 |
0.038 |
0.041 |
0.046 |
0.041 |
2 |
Weak positive serum
(N° 9350-07) |
1.072 |
1.053 |
0.943 |
1.017 |
1.021 |
46 |
Positive control serum
(N° 9350-08) |
2.215 |
2.301 |
2.049 |
2.280 |
2.211 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.211 |
Yes |
Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
46 % |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.110 |
Yes |
Comments:
Good for use.