#9350
Echinococcus
granulosus IgG ELISA
ELISA kit for the diagnosis of human echinococcoses
Lot 2411G (Expiry date: 28.10.2025)
Production date : 29.02.2024
Results of quality control 1 (after
production), date: 04.03.2024
No serum blank (average) : 0.140
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.022 |
-0.001 |
-0.008 |
0.030 |
0.000 |
0 |
Negative control serum
(N° 9350-06) |
0.024 |
0.079 |
0.029 |
0.125 |
0.064 |
3 |
Weak positive serum
(N° 9350-07) |
1.093 |
1.213 |
1.164 |
1.267 |
1.184 |
49 |
Positive control serum
(N° 9350-08) |
2.256 |
2.456 |
2.338 |
2.663 |
2.428 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9350-08 |
> 1.200 |
2.428 |
Yes |
Absorbance of weak positive serum 9350-07, as % of positive serum |
> 27 % |
49 % |
Yes |
Absorbance of negative serum 9350-06, as % of positive serum |
< 8 % |
3 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.140 |
Yes |
Comments:
Good for use.