| #9450
Strongyloïdes
ratti IgG ELISA ELISA kit for the diagnosis of human strongyloidosis
 Lot 2410S (Expiry date : 19.10.2025)
 Production date : 20.02.2024 Results of quality control 3(after
12 months), date: 13.02.2025 
 No serum blank (average) : 0.114
 
  
    | Sample | Absorbances of individual wells after blank reduction | Average | % pos. |  
    | No serum blank 
 | 0.000 | -0.001 | -0.001 | 0.002 | 0.000 | 0 |  
    | Negative control serum (N° 9450-06)
 | 0.049 | 0.043 | 0.035 | 0.047 | 0.044 | 2 |  
    | Weak positive serum (N° 9450-07)
 | 0.619 | 0.576 | 0.606 | 0.586 | 0.597 | 28 |  
    | Positive control serum (N° 9450-08)
 | 2.121 | 2.176 | 2.056 | 2.140 | 2.123 | 100 |  Validation criteria: 
  
    |  | Criterion | Measure | Valid |  
    | Absorbance of positive serum 9450-08 | > 1.200 | 2.123 | Yes |  
    | Absorbance of weak positive serum 9450-07, as % of positive
      serum | > 18 % | 28 % | Yes |  
    | Absorbance of negative serum 9450-06, as % of positive serum | < 10 % | 2 % | Yes |  
    | Absorbance of no serum blank | < 0.350 | 0.114 | Yes |  Comments:  Good for use.  Bordier Affinity Products SA, Chatanerie 2,