#9450 Strongyloïdes ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis

Lot 2410S (Expiry date : 19.10.2025)

Production date : 20.02.2024

Results of quality control 2 (after 6 months), date: 16.08.2024

No serum blank (average) : 0.111

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.004 -0.005 0.008 -0.007 0.000 0
Negative control serum
(N° 9450-06)
0.068 0.058 0.073 0.066 0.066 3
Weak positive serum
(N° 9450-07)
0.705 0.695 0.718 0.734 0.713 31
Positive control serum
(N° 9450-08)
2.241 2.311 2.292 2.453 2.324 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9450-08 > 1.200 2.324 Yes
Absorbance of weak positive serum 9450-07, as % of positive serum > 18 % 31 % Yes
Absorbance of negative serum 9450-06, as % of positive serum < 10 % 3 % Yes
Absorbance of no serum blank < 0.350 0.111 Yes

Comments:

Good for use.