#9450 Strongyloïdes ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis

Lot 2409S (Expiry date : 04.10.2025)

Production date : 05.02.2024

Results of quality control 3 (after 12 months), date: 30.01.2025

No serum blank (average) : 0.124

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.000 0.001 0.001 -0.002 0.000 0
Negative control serum
(N° 9450-06)
0.084 0.065 0.061 0.058 0.067 3
Weak positive serum
(N° 9450-07)
0.690 0.701 0.637 0.650 0.670 29
Positive control serum
(N° 9450-08)
2.214 2.510 2.207 2.216 2.287 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9450-08 > 1.200 2.287 Yes
Absorbance of weak positive serum 9450-07, as % of positive serum > 18 % 29 % Yes
Absorbance of negative serum 9450-06, as % of positive serum < 10 % 3 % Yes
Absorbance of no serum blank < 0.350 0.124 Yes

Comments:

Good for use.