#9450 Strongyloïdes ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis

Lot 2409S (Expiry date : 04.10.2025)

Production date : 05.02.2024

Results of quality control 2 (after 6 months), date: 06.08.2024

No serum blank (average) : 0.112

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.001 0.000 0.000 0.000 0.000 0
Negative control serum
(N° 9450-06)
0.079 0.078 0.077 0.088 0.080 4
Weak positive serum
(N° 9450-07)
0.678 0.679 0.699 0.710 0.691 31
Positive control serum
(N° 9450-08)
2.058 2.327 2.150 2.382 2.229 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9450-08 > 1.200 2.229 Yes
Absorbance of weak positive serum 9450-07, as % of positive serum > 18 % 31 % Yes
Absorbance of negative serum 9450-06, as % of positive serum < 10 % 4 % Yes
Absorbance of no serum blank < 0.350 0.112 Yes

Comments:

Good for use.