#9600 Schistosoma mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis

Lot 2408B (Expiry date: 14.09.2025)

 

Production date : 15.01.2024

Results of quality control 2 (after 6 months), date: 16.07.2024

No serum blank (average) : 0.152

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.004 -0.006 0.003 -0.002 0.000 0
Negative control serum
(N° 9600-06)
0.049 0.052 0.064 0.047 0.053 2
Weak positive serum
(N° 9600-07)
0.398 0.384 0.424 0.404 0.403 17
Positive control serum
(N° 9600-08)
2.240 2.468 2.299 2.391 2.350 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.350 Yes
Absorbance of weak positive serum 9600-07, as % of positive serum > 10 % 17 % Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.152 Yes

Comments:

Good for use.