Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9600 Schistosoma mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis

Lot 2408B (Expiry date: 14.09.2025)

 

Production date : 15.01.2024

Results of quality control 1 (after production), date: 23.01.2024

No serum blank (average) : 0.153

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.006 0.015 -0.010 0.001 0.000 0
Negative control serum
(N° 9600-06)
0.050 0.082 0.048 0.072 0.063 3
Weak positive serum
(N° 9600-07)
0.572 0.507 0.439 0.449 0.491 20
Positive control serum
(N° 9600-08)
2.466 2.409 2.365 2.558 2.449 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.449 Yes
Absorbance of weak positive serum 9600-07, as % of positive serum > 10 % 20 % Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.153 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch