Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9550 Entamoeba histolytica IgG ELISA
ELISA kit for the diagnosis of human visceral amebiasis

Lot 2406H (Expiry date: 14.09.2025)

 

Production date : 15.01.2024

Results of quality control 3 (after 12 months), date: 10.01.2025

No serum blank (average) : 0.154

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.006 -0.007 0.010 -0.009 0.000 0
Negative control serum
(N° 9550-06)
0.118 0.119 0.138 0.130 0.126 6
Weak positive serum
(N° 9550-07)
0.480 0.462 0.541 0.508 0.498 22
Positive control serum
(N° 9550-08)
2.087 2.226 2.291 2.322 2.232 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9550-08 > 1.200 2.232 Yes
Absorbance of weak positive serum 9550-07, as % of positive serum >  12 % 22 % Yes
Absorbance of negative serum 9550-06, as % of positive serum < 12 % 6 % Yes
Absorbance of no serum blank < 0.350 0.154 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch