#9400
Acanthocheilonema
viteae (Onchocercidae) IgG ELISA
ELISA kit for the diagnosis of human filariasis
Lot 2405V (Expiry date: 14.09.2025)
Production date : 15.01.2024
Results of quality control 4 (after
20 months), date: 15.09.2025
No serum blank (average) : 0.128
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.005 |
0.003 |
0.003 |
-0.001 |
0.000 |
0 |
Negative control serum
(N° 9400-06) |
0.114 |
0.106 |
0.099 |
0.090 |
0.102 |
4 |
Weak positive serum
(N° 9400-07) |
1.047 |
0.996 |
1.063 |
0.993 |
1.025 |
41 |
Positive control
serum
(N° 9400-08) |
2.525 |
2.490 |
2.443 |
2.502 |
2.490 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9400-08 |
> 1.200 |
2.490 |
Yes |
Absorbance of weak positive serum 9400-07, as % of positive serum |
> 19 % |
41 % |
Yes |
Absorbance of negative serum 9400-06, as % of positive serum |
< 10 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.128 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|