#9400 Acanthocheilonema viteae (Onchocercidae) IgG ELISA
ELISA kit for the diagnosis of human filariasis

Lot 2405V (Expiry date: 14.09.2025)

 

Production date : 15.01.2024

Results of quality control 3 (after 12 months), date: 10.01.2025

No serum blank (average) : 0.123

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.000 0.001 -0.001 0.000 0.000 0
Negative control serum
(N° 9400-06)
0.103 0.089 0.111 0.103 0.102 4
Weak positive serum
(N° 9400-07)
0.956 0.901 0.879 0.914 0.913 39
Positive control serum
(N° 9400-08)
2.244 2.416 2.262 2.340 2.316 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9400-08 > 1.200 2.316 Yes
Absorbance of weak positive serum 9400-07, as % of positive serum > 19 % 39 % Yes
Absorbance of negative serum 9400-06, as % of positive serum < 10 % 4 % Yes
Absorbance of no serum blank < 0.350 0.123 Yes

Comments:

Good for use.