#9400
Acanthocheilonema
viteae (Onchocercidae) IgG ELISA
ELISA kit for the diagnosis of human filariasis
Lot 2405V (Expiry date: 14.09.2025)
Production date : 15.01.2024
Results of quality control 3 (after
12 months), date: 10.01.2025
No serum blank (average) : 0.123
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.000 |
0.001 |
-0.001 |
0.000 |
0.000 |
0 |
Negative control serum
(N° 9400-06) |
0.103 |
0.089 |
0.111 |
0.103 |
0.102 |
4 |
Weak positive serum
(N° 9400-07) |
0.956 |
0.901 |
0.879 |
0.914 |
0.913 |
39 |
Positive control
serum
(N° 9400-08) |
2.244 |
2.416 |
2.262 |
2.340 |
2.316 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9400-08 |
> 1.200 |
2.316 |
Yes |
Absorbance of weak positive serum 9400-07, as % of positive serum |
> 19 % |
39 % |
Yes |
Absorbance of negative serum 9400-06, as % of positive serum |
< 10 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.123 |
Yes |
Comments:
Good for use.