#9400 Acanthocheilonema viteae (Onchocercidae) IgG ELISA
ELISA kit for the diagnosis of human filariasis

Lot 2405V (Expiry date: 14.09.2025)

 

Production date : 15.01.2024

Results of quality control 2 (after 6 months), date: 15.07.2024

No serum blank (average) : 0.112

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.002 -0.004 0.000 0.002 0.000 0
Negative control serum
(N° 9400-06)
0.099 0.094 0.106 0.098 0.099 4
Weak positive serum
(N° 9400-07)
0.992 0.962 0.952 0.950 0.964 42
Positive control serum
(N° 9400-08)
2.221 2.384 2.212 2.313 2.283 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9400-08 > 1.200 2.283 Yes
Absorbance of weak positive serum 9400-07, as % of positive serum > 19 % 42 % Yes
Absorbance of negative serum 9400-06, as % of positive serum < 10 % 4 % Yes
Absorbance of no serum blank < 0.350 0.112 Yes

Comments:

Good for use.