Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9400 Acanthocheilonema viteae (Onchocercidae) IgG ELISA
ELISA kit for the diagnosis of human filariasis

Lot 2405V (Expiry date: 14.09.2025)

 

Production date : 15.01.2024

Results of quality control 1 (after production), date: 23.01.2024

No serum blank (average) : 0.138

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.010 0.010 -0.015 0.014 0.000 0
Negative control serum
(N° 9400-06)
0.081 0.145 0.092 0.169 0.122 5
Weak positive serum
(N° 9400-07)
0.967 0.968 0.922 0.982 0.960 42
Positive control serum
(N° 9400-08)
2.175 2.372 2.178 2.477 2.301 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9400-08 > 1.200 2.301 Yes
Absorbance of weak positive serum 9400-07, as % of positive serum > 19 % 42 % Yes
Absorbance of negative serum 9400-06, as % of positive serum < 10 % 5 % Yes
Absorbance of no serum blank < 0.350 0.138 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch