Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9700 Taenia solium IgG ELISA
ELISA kit for the diagnosis of human cysticercosis

Lot 2404C (Expiry date: 07.09.2025)

Production date : 08.01.2024

Results of quality control 4 (after 20 months), date: 08.09.2025

No serum blank (average) : 0.198

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.002 -0.002 -0.002 0.001 0.000 0
Negative control serum
(N° 9700-06)
0.095 0.081 0.092 0.085 0.089 4
Weak positive serum
(N° 9700-07)
0.417 0.436 0.423 0.414 0.423 18
Positive control serum
(N° 9700-08)
2.332 2.458 2.301 2.475 2.392 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9700-08 > 1.200 2.392 Yes
Absorbance of weak positive serum 9700-07, as % of positive serum > 8 % 18 % Yes
Absorbance of negative serum 9700-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.198 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch