Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9700 Taenia solium IgG ELISA
ELISA kit for the diagnosis of human cysticercosis

Lot 2404C (Expiry date: 07.09.2025)

Production date : 08.01.2024

Results of quality control 1 (after production), date: 11.01.2024

No serum blank (average) : 0.226

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.007 0.010 -0.006 0.004 0.000 0
Negative control serum
(N° 9700-06)
0.069 0.128 0.086 0.112 0.099 4
Weak positive serum
(N° 9700-07)
0.422 0.478 0.407 0.431 0.434 19
Positive control serum
(N° 9700-08)
2.245 2.352 2.259 2.365 2.305 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9700-08 > 1.200 2.305 Yes
Absorbance of weak positive serum 9700-07, as % of positive serum > 8 % 19 % Yes
Absorbance of negative serum 9700-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.226 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch