Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9700 Taenia solium IgG ELISA
ELISA kit for the diagnosis of human cysticercosis

Lot 2403C (Expiry date: 07.09.2025)

Production date : 08.01.2024

Results of quality control 1 (after production), date: 11.01.2024

No serum blank (average) : 0.274

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.019 0.016 -0.006 0.009 0.000 0
Negative control serum
(N° 9700-06)
0.111 0.187 0.069 0.183 0.137 6
Weak positive serum
(N° 9700-07)
0.538 0.513 0.498 0.549 0.524 22
Positive control serum
(N° 9700-08)
2.223 2.379 2.282 2.468 2.338 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9700-08 > 1.200 2.338 Yes
Absorbance of weak positive serum 9700-07, as % of positive serum > 8 % 22 % Yes
Absorbance of negative serum 9700-06, as % of positive serum < 8 % 6 % Yes
Absorbance of no serum blank < 0.350 0.274 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland