#9700
Taenia solium IgG ELISA
ELISA kit for the diagnosis of human cysticercosis
Lot 2403C (Expiry date: 07.09.2025)
Production date : 08.01.2024
Results of quality control 1 (after
production), date: 11.01.2024
No serum blank (average) : 0.274
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.019 |
0.016 |
-0.006 |
0.009 |
0.000 |
0 |
Negative control serum
(N° 9700-06) |
0.111 |
0.187 |
0.069 |
0.183 |
0.137 |
6 |
Weak positive serum
(N° 9700-07) |
0.538 |
0.513 |
0.498 |
0.549 |
0.524 |
22 |
Positive control
serum
(N° 9700-08) |
2.223 |
2.379 |
2.282 |
2.468 |
2.338 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9700-08 |
> 1.200 |
2.338 |
Yes |
Absorbance of weak positive serum 9700-07, as % of positive serum
|
> 8 % |
22 % |
Yes
|
Absorbance of negative serum 9700-06, as % of positive serum |
< 8 % |
6 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.274 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|