#9200
Toxocara
canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis
Lot 2401T (Expiry date: 07.09.2025)
Production date : 08.01.2024
Results of quality control 3 (after
12 months), date: 03.01.2025
No serum blank (average) : 0.118
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.008 |
0.011 |
-0.005 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9200-06) |
0.002 |
0.006 |
0.010 |
0.016 |
0.009 |
0 |
Weak positive serum
(N° 9200-07) |
1.245 |
1.273 |
1.220 |
1.259 |
1.249 |
55 |
Positive control serum
(N° 9200-08) |
2.171 |
2.342 |
2.239 |
2.289 |
2.260 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9200-08 |
> 1.200 |
2.260 |
Yes |
Absorbance of weak positive serum 9200-07, as % of positive serum |
> 32 % |
55 % |
Yes |
Absorbance of negative serum 9200-06, as % of positive serum |
< 6 % |
0 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.118 |
Yes |
Comments:
Good for use.