#9200
Toxocara
canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis
Lot 2401T (Expiry date: 07.09.2025)
Production date : 08.01.2024
Results of quality control 2 (after
6 months), date: 08.07.2024
No serum blank (average) : 0.108
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.001 |
-0.001 |
0.003 |
0.000 |
0.000 |
0 |
Negative control serum
(N° 9200-06) |
0.014 |
0.024 |
0.029 |
0.019 |
0.021 |
1 |
Weak positive serum
(N° 9200-07) |
1.442 |
1.339 |
1.371 |
1.402 |
1.388 |
59 |
Positive control serum
(N° 9200-08) |
2.269 |
2.327 |
2.352 |
2.451 |
2.350 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9200-08 |
> 1.200 |
2.350 |
Yes |
Absorbance of weak positive serum 9200-07, as % of positive serum |
> 32 % |
59 % |
Yes |
Absorbance of negative serum 9200-06, as % of positive serum |
< 6 % |
1 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.108 |
Yes |
Comments:
Good for use.