#9200
Toxocara
canis IgG ELISA
ELISA kit for the diagnosis of human toxocarosis
Lot 2401T (Expiry date: 07.09.2025)
Production date : 08.01.2024
Results of quality control 1 (after
production), date: 12.01.2024
No serum blank (average) : 0.125
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.004 |
0.026 |
-0.012 |
-0.009 |
0.000 |
0 |
Negative control serum
(N° 9200-06) |
0.013 |
0.068 |
0.006 |
0.119 |
0.051 |
2 |
Weak positive serum
(N° 9200-07) |
1.406 |
1.316 |
1.319 |
1.231 |
1.318 |
60 |
Positive control serum
(N° 9200-08) |
2.190 |
2.331 |
2.042 |
2.188 |
2.188 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9200-08 |
> 1.200 |
2.188 |
Yes |
Absorbance of weak positive serum 9200-07, as % of positive serum |
> 32 % |
60 % |
Yes |
Absorbance of negative serum 9200-06, as % of positive serum |
< 6 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.125 |
Yes |
Comments:
Good for use.