#9750
Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis
Lot 2337R (Expiry date: 26.07.2025)
Production date : 27.11.2023
Results of quality control 3 (after
12 months), date: 21.11.2024
No serum blank (average) : 0.126
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.009 |
0.003 |
-0.007 |
0.013 |
0.000 |
0 |
Negative control serum
(N° 9750-06) |
0.035 |
0.033 |
0.033 |
0.035 |
0.034 |
2 |
Weak positive serum
(N° 9750-07) |
0.383 |
0.425 |
0.373 |
0.394 |
0.393 |
18 |
Positive control serum
(N° 9750-08) |
2.122 |
2.348 |
2.106 |
2.340 |
2.229 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9750-08 |
> 1.200 |
2.229 |
Yes |
Absorbance of weak positive serum 9750-07, as % of positive serum |
> 10 % |
18 % |
Yes |
Absorbance of negative serum 9750-06, as % of positive serum |
< 8 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.126 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bātiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|