Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9750 Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis

Lot 2337R (Expiry date: 26.07.2025)

 

Production date : 27.11.2023

Results of quality control 2 (after 6 months), date: 27.05.2024

No serum blank (average) : 0.121

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 -0.006 0.002 0.006 0.000 0
Negative control serum
(N° 9750-06)
0.057 0.047 0.056 0.057 0.055 2
Weak positive serum
(N° 9750-07)
0.411 0.409 0.422 0.400 0.411 17
Positive control serum
(N° 9750-08)
2.225 2.420 2.306 2.460 2.353 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9750-08 > 1.200 2.353 Yes
Absorbance of weak positive serum 9750-07, as % of positive serum >  10 % 17 % Yes
Absorbance of negative serum 9750-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.121 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch