Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9750 Trichinella spiralis IgG ELISA
ELISA kit for the diagnosis of human trichinellosis

Lot 2337R (Expiry date: 26.07.2025)

 

Production date : 27.11.2023

Results of quality control 1 (after production), date: 01.12.2023

No serum blank (average) : 0.117

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.001 -0.005 0.005 -0.001 0.000 0
Negative control serum
(N° 9750-06)
0.057 0.068 0.066 0.076 0.066 3
Weak positive serum
(N° 9750-07)
0.456 0.479 0.433 0.499 0.466 20
Positive control serum
(N° 9750-08)
2.204 2.422 2.177 2.610 2.353 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9750-08 > 1.200 2.353 Yes
Absorbance of weak positive serum 9750-07, as % of positive serum >  10 % 20 % Yes
Absorbance of negative serum 9750-06, as % of positive serum < 8 % 3 % Yes
Absorbance of no serum blank < 0.350 0.117 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bātiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch