#9600 Schistosoma mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis

Lot 2335B (Expiry date: 29.06.2025)

 

Production date : 30.10.2023

Results of quality control 3 (after 12 months), date: 25.10.2024

No serum blank (average) : 0.135

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.003 0.001 0.001 0.000 0.000 0
Negative control serum
(N° 9600-06)
0.059 0.050 0.063 0.050 0.056 2
Weak positive serum
(N° 9600-07)
0.426 0.462 0.444 0.428 0.440 19
Positive control serum
(N° 9600-08)
2.178 2.393 2.244 2.328 2.286 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9600-08 > 1.200 2.286 Yes
Absorbance of weak positive serum 9600-07, as % of positive serum > 10 % 19 % Yes
Absorbance of negative serum 9600-06, as % of positive serum < 8 % 2 % Yes
Absorbance of no serum blank < 0.350 0.135 Yes

Comments:

Good for use.