#9600
Schistosoma
mansoni IgG ELISA
ELISA kit for the diagnosis of human schistosomiasis
Lot 2335B (Expiry date: 29.06.2025)
Production date : 30.10.2023
Results of quality control 2 (after
6 months), date: 25.04.2024
No serum blank (average) : 0.136
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.000 |
-0.003 |
0.003 |
0.001 |
0.000 |
0 |
Negative control serum
(N° 9600-06) |
0.050 |
0.049 |
0.062 |
0.057 |
0.054 |
2 |
Weak positive serum
(N° 9600-07) |
0.398 |
0.406 |
0.364 |
0.413 |
0.395 |
18 |
Positive control serum
(N° 9600-08) |
2.120 |
2.250 |
2.165 |
2.276 |
2.203 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9600-08 |
> 1.200 |
2.203 |
Yes |
Absorbance of weak positive serum 9600-07, as % of positive serum |
> 10 % |
18 % |
Yes |
Absorbance of negative serum 9600-06, as % of positive serum |
< 8 % |
2 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.136 |
Yes |
Comments:
Good for use.