#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2334D (Expiry date: 29.06.2025)
Production date : 30.10.2023
Results of quality control 2 (after
6 months), date: 25.04.2024
No serum blank (average) : 0.126
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.004 |
0.000 |
0.002 |
0.002 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.093 |
0.086 |
0.100 |
0.107 |
0.097 |
4 |
Weak positive serum
(N° 9650-07) |
0.362 |
0.399 |
0.332 |
0.356 |
0.362 |
15 |
Positive control
serum
(N° 9650-08) |
2.321 |
2.421 |
2.334 |
2.629 |
2.426 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
2.426 |
Yes |
Absorbance of weak positive serum 9650-07, as % of positive serum |
> 8 % |
15 % |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 8 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.126 |
Yes |
Comments:
Good for use.