#9650 Fasciola hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 2334D (Expiry date: 29.06.2025)

 

Production date : 30.10.2023

Results of quality control 2 (after 6 months), date: 25.04.2024

No serum blank (average) : 0.126

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.004 0.000 0.002 0.002 0.000 0
Negative control serum
(N° 965
0-06)
0.093 0.086 0.100 0.107 0.097 4
Weak positive serum
(N° 965
0-07)
0.362 0.399 0.332 0.356 0.362 15
Positive control serum
(N° 9650
-08)
2.321 2.421 2.334 2.629 2.426 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.426 Yes
Absorbance of weak positive serum 9650-07, as % of positive serum > 8 % 15 % Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.126 Yes

Comments:

Good for use.