#9650 Fasciola hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis

Lot 2334D (Expiry date: 29.06.2025)

 

Production date : 30.10.2023

Results of quality control 1 (after production), date: 06.11.2023

No serum blank (average) : 0.124

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
0.000 -0.001 0.004 -0.003 0.000 0
Negative control serum
(N° 965
0-06)
0.048 0.088 0.072 0.128 0.084 4
Weak positive serum
(N° 965
0-07)
0.318 0.316 0.401 0.424 0.365 17
Positive control serum
(N° 9650
-08)
1.927 2.131 2.207 2.404 2.167 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9650-08 > 1.200 2.167 Yes
Absorbance of weak positive serum 9650-07, as % of positive serum > 8 % 17 % Yes
Absorbance of negative serum 9650-06, as % of positive serum < 8 % 4 % Yes
Absorbance of no serum blank < 0.350 0.124 Yes

Comments:

Good for use.