#9650
Fasciola
hepatica IgG ELISA (Trematoda)
ELISA kit for the diagnosis of human fascioliasis
Lot 2334D (Expiry date: 29.06.2025)
Production date : 30.10.2023
Results of quality control 1 (after
production), date: 06.11.2023
No serum blank (average) : 0.124
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
0.000 |
-0.001 |
0.004 |
-0.003 |
0.000 |
0 |
Negative control serum
(N° 9650-06) |
0.048 |
0.088 |
0.072 |
0.128 |
0.084 |
4 |
Weak positive serum
(N° 9650-07) |
0.318 |
0.316 |
0.401 |
0.424 |
0.365 |
17 |
Positive control
serum
(N° 9650-08) |
1.927 |
2.131 |
2.207 |
2.404 |
2.167 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9650-08 |
> 1.200 |
2.167 |
Yes |
Absorbance of weak positive serum 9650-07, as % of positive serum |
> 8 % |
17 % |
Yes |
Absorbance of negative serum 9650-06, as % of positive serum |
< 8 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.124 |
Yes |
Comments:
Good for use.