#9450
Strongyloïdes
ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis
Lot 2332S (Expiry date : 02.06.2025)
Production date : 03.10.2023
Results of quality control 3 (after
12 months), date: 26.09.2024
No serum blank (average) : 0.123
Sample |
Absorbances of individual wells after blank reduction |
Average |
% pos. |
No serum blank
|
-0.002 |
0.001 |
0.001 |
0.000 |
0.000 |
0 |
Negative control serum
(N° 9450-06) |
0.080 |
0.068 |
0.079 |
0.074 |
0.075 |
4 |
Weak positive serum
(N° 9450-07) |
0.704 |
0.717 |
0.695 |
0.714 |
0.708 |
34 |
Positive control serum
(N° 9450-08) |
2.014 |
2.247 |
1.999 |
2.051 |
2.078 |
100 |
Validation criteria:
|
Criterion |
Measure |
Valid |
Absorbance of positive serum 9450-08 |
> 1.200 |
2.078 |
Yes |
Absorbance of weak positive serum 9450-07, as % of positive
serum |
> 18 % |
34 % |
Yes |
Absorbance of negative serum 9450-06, as % of positive serum |
< 10 % |
4 % |
Yes |
Absorbance of no serum blank |
< 0.350 |
0.123 |
Yes |
Comments:
Good for use.
Bordier Affinity Products SA, Chatanerie 2,
Bâtiment Biokema,
CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67. Fax: +
41 21 633 31 78. Email: cb(at)bordier.ch
|