Home    Products    Sales Conditions & distributors    Quality controls    Instructions For Use    Links

#9450 Strongyloïdes ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis

Lot 2332S (Expiry date : 02.06.2025)

Production date : 03.10.2023

Results of quality control 3 (after 12 months), date: 26.09.2024

No serum blank (average) : 0.123

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.002 0.001 0.001 0.000 0.000 0
Negative control serum
(N° 9450-06)
0.080 0.068 0.079 0.074 0.075 4
Weak positive serum
(N° 9450-07)
0.704 0.717 0.695 0.714 0.708 34
Positive control serum
(N° 9450-08)
2.014 2.247 1.999 2.051 2.078 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9450-08 > 1.200 2.078 Yes
Absorbance of weak positive serum 9450-07, as % of positive serum > 18 % 34 % Yes
Absorbance of negative serum 9450-06, as % of positive serum < 10 % 4 % Yes
Absorbance of no serum blank < 0.350 0.123 Yes

Comments:

Good for use.

Bordier Affinity Products SA, Chatanerie 2, Bâtiment Biokema, CH-1023 Crissier, Switzerland
Phone: + 41 21 633 31 67.  Fax: + 41 21 633 31 78. Email: cb(at)bordier.ch