#9450 Strongyloïdes ratti IgG ELISA
ELISA kit for the diagnosis of human strongyloidosis

Lot 2332S (Expiry date : 02.06.2025)

Production date : 03.10.2023

Results of quality control 2 (after 6 months), date: 28.03.2024

No serum blank (average) : 0.137

Sample

Absorbances of individual wells after blank reduction

Average

% pos.

No serum blank
-0.020 0.017 -0.009 0.012 0.000 0
Negative control serum
(N° 9450-06)
0.084 0.116 0.071 0.115 0.097 5
Weak positive serum
(N° 9450-07)
0.624 0.658 0.673 0.672 0.657 33
Positive control serum
(N° 9450-08)
1.975 1.993 1.995 2.040 2.001 100

Validation criteria:

  Criterion Measure Valid
Absorbance of positive serum 9450-08 > 1.200 2.001 Yes
Absorbance of weak positive serum 9450-07, as % of positive serum > 18 % 33 % Yes
Absorbance of negative serum 9450-06, as % of positive serum < 10 % 5 % Yes
Absorbance of no serum blank < 0.350 0.137 Yes

Comments:

Good for use.